ΤΟ ΑΡΘΡΟ ΤΟΥ ΤΜΗΜΑΤΟΣ ΕΝΔΟΟΥΡΟΛΟΓΙΑΣ - ΛΑΠΑΡΟΣΚΟΠΙΚΗΣ Χ/Ρ ΟΥΡΟΛΟΓΙΑΣ & ΟΥΡΟΛΟΓΙΚΗΣ ΤΕΧΝΟΛΟΓΙΑΣ
Πίνακας 3, αναπροσαρμογή από Panthier et al. [5]
Study
year
Experiment
Pressure Measurement
Results
Bhojani et al. [18]
2023
Retrospective observational analysis
Retrogradely, pressure sensor on LithoVueTM Elite uretersocope
50 cases Τhe median IRP was 28.5 (20.0–47.5) mmHg and the max. IRP was 174.0 (133.5– 266.0) mmHg. IRP remained at <60 mmHg during 92% of the procedure times.
Sierra et al. [19]
2023
Prospective cohort study
Retrogradely (tip)
20 cases of fURS with 40 cmH2O irrigation Baseline IRP was 13.6 (6.8--47.6) cmH2O After UAS placement, baseline IRP was 17 (8--44.6) cmH2O. With 40 cmH2O irrigation, median IRP was 34 (19--81.6) cmH2O. During lithotripsy, max IRP recorded
Gao et al. [17]
2022
Retrospective cohort study Efficiency and safety
Pressure sensor at the UAS tip with suction
310 cases One session and 1-month SFRs were 80.65% and 82.26%. Clavien-Dindo grade II--III complication was 3.26%
Antonucci et al. [22]
2022
Retrospective cohort study Infectious risk
No direct pressure measurement Standardized technique
of gravity-based
irrigation and opened-5Fr aside ureteral catheter
236 Patients Sepsis occurred in 10 (4.2%) patients and septic shock occurred in 1 (0.4%) history of UTI predicted for higher risk of postoperative infections (OR 8.434, CI 95% 2.36--29.46). No difference in postoperative infections between expert and young surgeons
Patel et al. [9]
2021
Prospective cohort study
Retrogradely, through URS working channel (tip)
8 patients Lower intracalyceal pressure with UAS Compared to patients with a 12/14Fr UAS, those with a 14/16Fr UAS had significantly lower pressure in the interpolar (25.3 ± 13.1 vs. 44.0 ± 27.5 mmHg, P 1⁄4 0.03) and lower pole (16.2 ± 3.5 vs. 49.2 ± 40.3 mmHg, P 1⁄4 0.004) calyces
Doizi et al. [13]
2021
Prospective cohort study
Retrogradely (tip)
4 patients URS in the renal cavities and irrigation pressure set at 80 cmH2O without any forced irrigation, the mean
IRP was 63 cmH2O During laser lithotripsy, mean IRP with the use of ondemand forced irrigation was 115.3 cmH2O. Maximum
IRP peaks recorded during therapeutic period using forced irrigation
Έχουν εφαρμοσθεί διάφοροι τρόποι καταγραφής της ενδονεφρικής πίεσης χωρίς όμως μέχρι σήμερα να υπάρχει ένας κοινά αποδεκτός αξιόπιστος τρόπος μέτρησης [5]. Στις περισσότερες μελέτες οι μετρήσεις πραγματοποιήθηκαν σε πειραματικές συνθήκες, in vitro σε συνθετικά πειραματικά μοντέλα αλλά και
ex vivo σε ανθρώπινα και ζωικά (χοιρίδια) όργανα. Τα τελευταία χρόνια έχουν παρουσιασθεί κλινικές μελέτες παρατήρησης της ενδονεφρικής πίεσης σε ασθενείς, σε πραγματικές συνθήκες χειρουργείου, από τις οποίες προκύπτουν ενδιαφέροντα αποτε- λέσματα (Πίνακας 3) [5].
ΚΛΙΝΙΚΕΣ ΜΕΛΕΤΕΣ ΜΕ ΜΕΤΡΗΣΗ ΕΝΔΟΝΕΦΡΙΚΗΣ ΠΙΕΣΗΣ IRP
TTεεύύχχοοςς2288|| Mάρτιος - Απρίλιος 2024
Jung H, et al. [4]
65